You signed in with another tab or window. Reload to refresh your session.You signed out in another tab or window. Reload to refresh your session.You switched accounts on another tab or window. Reload to refresh your session.Dismiss alert
Hello sir
Thank you for your work sir, I have used your code sample sir.
I carried out impedance measurements with Rcal 10kohm and the results showed an accurate 10kohm, but when I took measurements on different resistors the results were far from accurate.
my example measured a 200kohm resistor but the results showed around 20kohm.
so how to solve this sir?
The text was updated successfully, but these errors were encountered:
Depending on the setting you have used for the PGA-gain, excitation voltage range and value for Rfb, your limits of measurement (i.e min/max impedance) may vary. Could you please specify these parameters?
Hello sir
Thank you for your work sir, I have used your code sample sir.
I carried out impedance measurements with Rcal 10kohm and the results showed an accurate 10kohm, but when I took measurements on different resistors the results were far from accurate.
my example measured a 200kohm resistor but the results showed around 20kohm.
so how to solve this sir?
The text was updated successfully, but these errors were encountered: